• Eight-channel laser diode portable airborne particle counter designed for ISO Class 1-7 cleanrooms
• Pass/fail report wizard for ISO-14644-1, EU GMP Annex 1, and FS-209E report formats
• Connect Solair model to a manifold for real-time cleanroom particle contamination monitoring
• Includes: Operating manual with CD, LMS Express transfer software download instructions, isokinetic sample probe, tripod, tubing, purge filter, power cables, spare fuse, printer, printer paper
• Additional 8-ch sizes available
• Applications: cleanroom certification, troubleshooting, routine testing, compressed gas testing, HEPA filter scanning, HEPA filter challenge testing, environmental testing
• Meets ISO 21501-4 calibration
• Zero Count Level<1 count / 5 minutes (per JIS B 9921) on 3100, 3350, 5100 and Boulder models
• Flow Rate: 1.0 CFM (28.3 LPM)
• Calibration Meets ISO 21501-4 calibration using NIST traceable PSL spheres, DMA and condensation Particle Counter Security: Integrated alarm buzzer and external alarm connector, alarms on counts, low battery, sensor failure
• Sample Output: Internally filtered to 100% @ 0.1 μm, optional HEPA filtration for cooling system exhaust
• Features: stainless steel enclosure; icon-driven 5.7 - inch (14.47 cm) color TFT touch screen display; USB, Ethernet, RS-485, Secure Data Transfer to and Storage on USB Flash Drive; automatic, manual, beep, concentration, cumulative/differential count modes; rotating buffer (includes particle and environmental data, location and time); Languages English (U.S.), French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean
• Vacuum Source: Internal closed-loop controlled pump. Allows the use of 3/8" ID sample tubing of up to 12 meters (40 feet) @ 50 & 100L
• Optional Environmental Sensors(up to 4) temperature/relative humidity, air velocity, differential pressure or others
• Supporting Software (sold separately): LMS Express, LMS Express RT, LMS Pharma
• Optional channels available on all Solair Airborne Particle Counter models